Verifying a liquid handler with dye or gravimetry

Last reviewed Author: BioStudioAI

Answer first. A liquid handler is trusted only after measured verification: dye absorbance or weighing across the deck, per channel and volume.

A liquid handler is trusted only after measured verification: dye absorbance or weighing across the deck, per channel and volume.

Principle

Vendor-neutral workflow; follow kit inserts for kit volumes and validate in your matrix.

Materials and equipment

ItemNotes
Samples, standards, controlsEvery run, lots recorded
ReagentsGeneric names, supplier/lot recorded
EquipmentCalibrated pipettes, suitable instrument

Procedure

  1. Run the verification protocol (dye or gravimetric) for each volume used.
  2. Compute per-channel accuracy and CV.
  3. Service or adjust channels outside limits; document the result.
  4. Repeat on the lab's schedule and after maintenance.

Expected results and QC

All channels within the lab's stated accuracy/CV limits.

Data analysis

Matching Tools calculator or Guide; raw file kept, settings recorded.

Safety and waste

Dyes and solvents: gloves. Research use only.

Frequently asked questions

First check?

Controls versus the last good run.

References

  1. Assay Guidance Manual · NCBI Bookshelf (2024)
  2. MIQE Guidelines · Clinical Chemistry (2009)
  3. Cold Spring Harbor Protocols · CSH Press (2024)

Do this analysis in BioStudioAI

Upload the raw file and BioStudioAI fits, quality-checks and reports it with an audit trail.

See the platform →

Related

Miniaturizing an assay from 96 to 384 wells

Volumes shrink 4-fold but evaporation, meniscus and pipetting errors do not. Re-validate Z-prime, edge behaviour and liquid handling at the new scale

The precision profile: CV across the range

Precision varies with concentration, worst near the limits. A precision profile (CV versus level) defines where the assay may honestly report numbers

Transferring an assay between labs or sites

Assay transfer is a designed comparison: shared samples and standards, pre-set equivalence criteria, and both sites' raw data retained. A protocol PDF

Plate assay QC